B875–111 Analog Input, Calibration
(Original Document)
Calibration Tools
The analog input module is calibrated at the factory prior to shipment. To ensure the module’s accuracy, the trim-pot should be calibrated regularly on a yearly interval.
The following tools and materials are needed to calibrate an analog input module:
Calibration Procedure
Use the following table to calibrate the module.
Step
Action
1
Remove two screws and label located on front-panel of analog input module to allow access to trim pot.
Note: The trim pot adjusts the total magnitude of the range (full scale). Refer to trim-pot location diagram, below.
2
Open analog input module handle to expose connectors and terminals.
3
Connect digital voltmeter (DVM) minus lead to voltage reference - (terminal number 38) and plus lead to voltage reference + (terminal number 39). Refer to voltmeter connection diagram, below.
Note: Field wiring may remain connected during calibration. Five minutes of warm-up is sufficient to attain temperature stability.
4
Adjust reference voltage trim pot for an indication of 4.500 V 0.0001 V.
5
Secure trim pot in place with adhesive.
6
Disconnect voltmeter, return connections to their pre-calibration state, close module handle. Replace front label removed in step 1, above.
Trim Pot Location Diagram
Trim pot location diagram.
Voltmeter Connection Diagram.
Voltmeter connection diagram.