Temperature
|
Thermocouple B
|
Thermocouple E
|
Thermocouple J
|
Thermocouple K
|
||||
---|---|---|---|---|---|---|---|---|
Conversion resolution (1)
|
0.24°C
|
0.026°C
|
0.037°C
|
0.048°C
|
||||
View resolution
|
0.1°C
|
0.1°C
|
0.1°C
|
0.1°C
|
||||
Maximum error at 25°C (2)
|
IC / EC (3)
|
IC
|
EC
|
IC
|
EC
|
IC
|
EC
|
|
Operating point
|
-200°C
|
16.8°C
|
2.7°C
|
18.7°C
|
3.3°C
|
|||
-100°C
|
9.5°C
|
1.7°C
|
9.5°C
|
1.8°C
|
||||
0°C
|
7.5°C
|
1.5°C
|
7.4°C
|
1.5°C
|
7.5°C
|
1.6°C
|
||
100°C
|
6.7°C
|
1.4°C
|
7.1°C
|
1.5°C
|
7.4°C
|
1.7°C
|
||
200°C
|
6.2°C
|
1.5°C
|
7.1°C
|
1.7°C
|
7.8°C
|
1.9°C
|
||
300°C
|
6.1°C
|
1.5°C
|
7.3°C
|
1.8°C
|
7.6°C
|
2.0°C
|
||
400°C
|
6.1°C
|
1.7°C
|
7.4°C
|
2.0°C
|
7.6°C
|
2.1°C
|
||
500°C
|
6.2°C
|
1.8°C
|
7.5°C
|
2.1°C
|
7.8°C
|
2.3°C
|
||
600°C
|
4.7°C
|
6.4°C
|
2.0°C
|
7.3°C
|
2.2°C
|
7.9°C
|
2.4°C
|
|
700°C
|
4.0°C
|
6.6°C
|
2.1°C
|
7.0°C
|
2.2°C
|
8.2°C
|
2.6°C
|
|
800°C
|
4.0°C
|
6.8°C
|
2.3°C
|
8.6°C
|
2.8°C
|
|||
900°C
|
3.8°C
|
8.9°C
|
3.1°C
|
|||||
1,000°C
|
3.6°C
|
9.3°C
|
3.3°C
|
|||||
1,100°C
|
3.5°C
|
9.8°C
|
3.6°C
|
|||||
1,200°C
|
3.6°C
|
10.3°C
|
3.8°C
|
|||||
1,300°C
|
3.6°C
|
|||||||
1,400°C
|
3.5°C
|
|||||||
1,500°C
|
3.5°C
|
|||||||
1,600°C
|
3.7°C
|
|||||||
1,700°C
|
3.9°C
|
|||||||
Input dynamic (4)
|
0..1802°C
|
-270..812°C
|
-210..1065°C
|
-270..1372°C
|
||||
Legend:
|
||||||||
(1)
|
These values appear in the middle of the thermocouple range.
|
|||||||
(2)
|
IC: ambient temperature of the TSX AEY 414 (20°C) and automatic internal compensation.
EC: ambient temperature of the TSX AEY 414 (30°C) and class A Pt100 automatic external compensation.
|
|||||||
(3)
|
With thermocouple B, the type of cold junction compensation (internal or external) is not taken into account, as this has no effect on precision.
|
|||||||
(4)
|
Internal compensation: ambient temperature = 20 °C
External compensation: ambient temperature = 30 °C.
|
Temperature
|
Thermocouple L
|
Thermocouple N
|
Thermocouple R
|
Thermocouple S
|
|||||
---|---|---|---|---|---|---|---|---|---|
Conversion resolution (1)
|
0.036°C
|
0.05°C
|
0.16°C
|
0.19°C
|
|||||
View resolution
|
0.1°C
|
0.1°C
|
0.1°C
|
0.1°C
|
|||||
Maximum error at 25°C (2)
|
IC
|
EC
|
IC
|
EC
|
IC
|
EC
|
IC
|
EC
|
|
Operating point
|
-200°C
|
19.6°C
|
4.0°C
|
||||||
-100°C
|
9.5°C
|
2.1°C
|
|||||||
0°C
|
7.5°C
|
1.5°C
|
7.8°C
|
1.8°C
|
11.4°C
|
4.8°C
|
11.2°C
|
4.7°C
|
|
100°C
|
7.1°C
|
1.5°C
|
7.0°C
|
1.8°C
|
8.1°C
|
3.5°C
|
8.3°C
|
3.5°C
|
|
200°C
|
7.2°C
|
1.7°C
|
6.5°C
|
1.7°C
|
7.1°C
|
3.2°C
|
7.4°C
|
3.3°C
|
|
300°C
|
7.3°C
|
1.9°C
|
6.2°C
|
1.8°C
|
6.5°C
|
2.9°C
|
6.9°C
|
3.1°C
|
|
400°C
|
7.5°C
|
2.0°C
|
6.0°C
|
1.9°C
|
6.3°C
|
3.0°C
|
6.8°C
|
3.2°C
|
|
500°C
|
7.4°C
|
2.1°C
|
6.0°C
|
2.0°C
|
6.2°C
|
3.0°C
|
6.8°C
|
3.3°C
|
|
600°C
|
7.4°C
|
2.2°C
|
6.1°C
|
2.1°C
|
6.1°C
|
3.1°C
|
6.8°C
|
3.4°C
|
|
700°C
|
7.1°C
|
2.2°C
|
6.2°C
|
2.2°C
|
6.1°C
|
3.1°C
|
6.6°C
|
3.3°C
|
|
800°C
|
6.8°C
|
2.3°C
|
6.3°C
|
2.4°C
|
6.0°C
|
3.2°C
|
6.6°C
|
3.4°C
|
|
900°C
|
6.7°C
|
2.3°C
|
6.5°C
|
2.6°C
|
6.0°C
|
3.2°C
|
6.6°C
|
3.5°C
|
|
1,000°C
|
6.8°C
|
2.7°C
|
5.9°C
|
3.3°C
|
6.6°C
|
3.6°C
|
|||
1,100°C
|
7.0°C
|
2.9°C
|
5.9°C
|
3.3°C
|
6.6°C
|
3.7°C
|
|||
1,200°C
|
7.4°C
|
3.2°C
|
5.9°C
|
3.4°C
|
6.7°C
|
3.8°C
|
|||
1,300°C
|
6.0°C
|
3.5°C
|
6.8°C
|
3.9°C
|
|||||
1,400°C
|
6.1°C
|
3.7°C
|
6.9°C
|
4.1°C
|
|||||
1,500°C
|
6.3°C
|
3.8°C
|
7.2°C
|
4.3°C
|
|||||
1,600°C
|
6.5°C
|
4.0°C
|
7.5°C
|
4.5°C
|
|||||
Input dynamic (3)
|
-200..900°C
|
-270..1300°C
|
-50..1769°C
|
-50..1769°C
|
|||||
Legend:
|
|||||||||
(1)
|
These values appear in the middle of the thermocouple range.
|
||||||||
(2)
|
IC: ambient temperature of the TSX AEY 414 (20°C) and automatic internal compensation.
EC: ambient temperature of the TSX AEY 414 (30°C) and class A Pt100 automatic external compensation.
|
||||||||
(3)
|
Internal compensation: ambient temperature = 20 °C,
External compensation: ambient temperature = 30 °C.
|
Temperature
|
Thermocouple T
|
Thermocouple U
|
|||
---|---|---|---|---|---|
Conversion resolution (1)
|
0.046°C
|
0.038°C
|
|||
View resolution
|
0.1°C
|
0.1°C
|
|||
Maximum error at 25°C (2)
|
IC
|
EC
|
IC
|
EC
|
|
Operating point
|
-200°C
|
18.3°C
|
3.2°C
|
||
-150°C
|
13.0°C
|
2.4°C
|
|||
-100°C
|
10.3°C
|
2.0°C
|
|||
-50°C
|
8.7°C
|
1.7°C
|
|||
0°C
|
7.7°C
|
1.6°C
|
7.7°C
|
1.6°C
|
|
50°C
|
7.1°C
|
1.5°C
|
|||
100°C
|
6.6°C
|
1.5°C
|
6.7°C
|
1.5°C
|
|
150°C
|
6.2°C
|
1.5°C
|
|||
200°C
|
5.9°C
|
1.5°C
|
5.8°C
|
1.5°C
|
|
250°C
|
5.7°C
|
1.5°C
|
|||
300°C
|
5.6°C
|
1.5°C
|
5.4°C
|
1.5°C
|
|
350°C
|
5.5°C
|
1.6°C
|
|||
400°C
|
5.4°C
|
1.6°C
|
|||
500°C
|
5.2°C
|
1.6°C
|
|||
600°C
|
5.0°C
|
1.7°C
|
|||
Input dynamic (3)
|
-270..400°C
|
-200..600°C
|
|||
Legend:
|
|||||
(1)
|
These values appear in the middle of the thermocouple range.
|
||||
(2)
|
IC: ambient temperature of the TSX AEY 414 (20°C) and automatic internal compensation.
EC: ambient temperature of the TSX AEY 414 (30°C) and class A Pt100 automatic external compensation.
|
||||
(3)
|
Internal compensation: ambient temperature = 20 °C,
External compensation: ambient temperature = 30 °C.
|